KYORITSU 2055 AC/DC Digital Clamp Meters (KEW 2055)
Couldn't load pickup availability
Got questions?
Feel free to WhatsApp us
Product Brochure
Click Here: Brochure in PDF
Product Datasheet / User Manual
Click Here: Datasheet in PDF
KEW 2055
Designed to meet international safety standard IEC 61010-1 CAT.IV 600V.
CAT.IV Clamp Meters can measure the Voltage and Current in both very low and high power circuits.
Thus, very useful for power distribution companies, power utilities and maintenance fields.
AC A
0~600.0/1000A
|±1.5%rdg±5dgt (50/60Hz)
|±3.0%rdg±5dgt (40~400Hz)
DC A
0~600.0/1000A
|±1.5%rdg±5dgt
AC V
6/60/600V
Auto Ranging
|±1.3%rdg±4dgt (50/60Hz)
|±3.0%rdg±5dgt (40~400Hz)
DC V
600m/6/60/600V
Auto Ranging
|±1.0%rdg±3dgt
Ω
600/6k/60k/600k/6M/60MΩ (Auto Ranging)
|±1%rdg±5dgt (600~6M) / ±5%rdg±8dgt (60M)
Continuity buzzer
Buzzer Sounds at 100Ω
Frequency/DUTY
10/100/1k/10kHz (Auto Ranging) / 0.1~99.9%
Conductor Size : Ø40mm
Withstand Voltage : 6880V AC for 5 seconds
Safety Standard
IEC 61010-1 CAT.IV 600V,
IEC 61010-031,
IEC 61010-2-032,
IEC 61326
Power Source : R03 (1.5V) (AAA) × 2
Dimensions
254(L) × 82(W) × 36(D)mm (26.5 x 14 x 5.6 cm)
Weight : 310g
Accessories
7066A (Test leads)
9094 (Carrying case)
Instruction Manual
Optional
8008 (Multi-Tran)
Made in: Thailand